Standard Environment
Temperature: 22 ℃ - 28 ℃
Relative humidity: 25% RH - 85% RH
Air pressure: 86 kPa - 106 kPa
Dark-room Conditions
The laboratory background brightness contribution for measuring spatial reflection must be less than 1/20 of the minimum brightness output of the device under test (DUT). If this standard is not met, background brightness deduction must be performed and recorded in the report.
Pre-work
The measurement should be conducted under standard conditions after sufficient preheating, with the equipment preheated for 15 minutes upon startup. The test sample should be tested after the lighting source and DUT have been fully preheated, as the brightness output of the DUT takes time to stabilize. Test the sample in a preheated stable state. During the measurement period, all measurement conditions should be kept constant.